Hårdhetsmätare Duroline M2

ModelM2
Vickers HV as per ISO 6507, ASTM E384, E92HV0,01 - HV0,025 - HV0,05 - HV0,1- HV0,2 - HV0,3 - HV0,5 - HV1
Knoop HK as per ISO 4545, ASTM E384, E92HK0,01 - HK0,025 - HK0,05 - HK0,1- HK0,2 - HK0,3 - HK0,5 - HK1
Hardness conversions to other scalesAs per ISO 18265
Test Loads (gf)10gf – 25gf – 50gf – 100gf – 200gf - 300gf- 500gf - 1000gf
Test Loads (N)0,09807N - 0,2452N - 0,4904N - 0,9807N - 1,961N - 2,942N - 4,904N - 9,807N
Test Load Accuracy10gf - 200gf: +/- 1,5% and > 200gf: +/- 1,0%
As per ISO standard (and ASTM E-384, JIS B7725, JIS B7734)
Test Load Dwell Time from 5 - 99 sec with 1 sec increments
As per ISO standard (and ASTM E-384, JIS B7725, JIS B7734)
Test Load Application Speed10gf - 200gf: < 0.070 mm/s and > 200gf: < 0.200 mm/s
As per ISO standard (and ASTM E-384, JIS B7725, JIS B7734)
Load Force ApplicationDead-weight, Automatic motorized, load, dwell, unload
Load SelectionManual by external setting knob
Turret ConfigurationMotorized 3-mount turret: 1 indenter, 2 objectives Position Auto-indication
IndenterIncluding 1 Vickers Diamond Indenter (ISO/DaKKs certified, Made in Germany)
ObjectivesStandard objectives 10X, 40X, optional 5X, 20X (Made in Germany)
Image capture by integrated CCD-camera (no eyepiece)
Sample Image CaptureBy Integrated CCD Camera, 5.0 MP, USB 3.0
(Made in Germany)
Indent Measurement Resolution0,00002 mm (0,02 µm)
Illumination SourceBuilt-in LED Light Source
Automatic Brightness ControlStandard for highest contrast and sharpness
Z-axis (Spindle Vertical Movement)Motorized